Prelab
Questions: Questions
to be answered before doing the experiment. The
answers are
due at the beginning of each
experiment without exception (the questions
are for credit and some may appear on your
final exam).
Compare MS ion trap
and electrospray ionization (ESI) technology to
the GC- MS quatrapole detection system.
Why does the LC-MS
tend to be the instrument of choice for many
biological samples?
Describe the
components of the Varian (Agilent) LC-MS (MCAL
lab) and how it is used in sample
identification.You can follow the sample through the MS
and describe what is happening at each step.
What is meant by
MS-MS, and what are the advantages of an
instrument that has this capability?
Objectives:
Learn the principles
and operation of LC-MS.
Develop methodology
using MS and MS-MS in sample identification.
Become familiar with
the concept of product ion fragmentation
pathways.
Apply methodology to
sample analysis.
Introduction:
The
MS in MCAL uses electro-spray
ionization (ESI) and Ion Trap technology to
detect sample ions. The analyte
ions are nebulized and dried (and become charged
ions).The
ions can be scanned in full scale mode,
MS-MS mode or can be further fragmented and the
product ions scanned.
The
MS is connected to a binary
LC system. Depending
on the column used
the LC can separate components in a mixture
based on hydrophobicity, charge, and
molecular size. A
major difference
between traditional HPLC and the chromatography
used in LC-MS is that in the
latter case the scale is usually much smaller,
both with respect to the
internal diameter of the column and even more so
with respect to flow
rate.The
MS will give good sensitivity
at flow rates of 200 μL/min or less. There are a
lot of other mass analyzers besides
Ion Trap that can be used in LC-MS; Single
Quadrupole, Triple Quadrupole, TOF (Time
of Flight) and Quadrupole-Time of Flight
(Q-TOF).
Reagents and Standards:
1.Milli-Q
water
2.C8 RP
column
3.HPLC-MS
acetonitrile (Fisher gold) formic acid
4.Syringe
filters
5.Analytical
balance
6.Pipettes
and disposable tips
7.Volumetric
flasks
8.Standards:
Acetaminophen,
acetylsalicylic acid, caffeine, and ibuprofen
9.Analgesic
medication of unknown concentration.
Laboratory
week 1:
optimization.
Standards of acetaminophen,acetylsalicylic acid,
caffeine, and ibuprofen
are supplied to the students. Prepare a
10
ug/ml solution of each standard using the
stocks and diluting solution provided in the 1.5
mL centrifuge tubes.1.5 mL will be sufficient.
Filter all solutions through a
0.45 um syringe filter.Fill the
MS syringe pump apparatus with the
filtered solution, connect to ESI unit with
syringe connector and peek tubing and
infuse the solution into the MS by turning on
syringe pump. Purge for
a few seconds (injects 200
ul/min) and leave on run
(which
injects 20 ul/min of sample).
The ion source, detector and RF
should be on.Click on them if they are off. They will be
green and show larger print when on.
The MS should be set to full
scan mode which is in the active
segment window.Set the
range from 105
to 500 so the instrument is not producing a lot of
excess data.Monitor the ions in the left hand bottom
screen and identify the major ion(s) of interest.Your ion
will be the M+H ion (MW plus 1 H
ion). If the major ion, that correspond to an
ingredient in your sample, is not
present it may be present as the M + Na (+23) ion
or as a breakdown product ion.
For each of acetaminophen,acetylsalicylic acid,
caffeine, and ibuprofen
optimize the needle voltage, capillary voltage and
% RF loading in the MS on
the major ions independently.This can
be done under the optimization
plots
screen, inputting the molar mass of the ion to
optimize in plot these masses window, and
selecting the parameter to optimize in
dependant
parameter window. The MS
will optimize each of needle voltage, capillary
voltage and %RF as you select
them and pressing start
icon.
Set the instrument in MS-MS
mode (under active segment), set the
optimization parameters (just determined)
for each of the major ions of interest, one at a
time and set the CID voltage
to 0 and you will see on the screen the parent ion
of interest.
To see the product ions
produced from each precursor ion; fragment
the ions using the MS-MS mode. To do this you must
determine the optimum CID
(excitation voltage) voltages to produce product
ions.Choose
method development (AMD) and
adjust the CID voltage (the
table allows you to use increasing values in
different channels) and monitor product
ion production in the different channels.
Once the optimum CID voltages
have been determined (usually
you want to see a little precursor ion) these can
be set in the MS-MS mode
window and product ion production monitored.This can be done for each ion of interest.Collect
the spectral data of the precursor and
product ions after optimization.
Repeat the MS-MS 3
and follow all product
ions.Make
a diagram of product ions
breakdown (formation) and demonstrate in lab.
write up how they are connected (fractionated)
to precursor ion.Try and identify as
many of the major product ions (molecular formula)
as you are able.
Collect all your data (your
data should go into a folder as
a descriptive name of run under 3590, 2015,
Group). Your chromatograms can be
copied into word and from word copied into your
write up.
HPLC
Seperation:
Check to
see how the elution
solvent(s), HPLC pumps, solvent mixer, auto-sample
and injector, column (C8 reverse
phase), and ESI unit are connected.
Start up HPLC pumps making sure
they are primed and are
pumping solvent.Using the windows
screen open the pump control screen (212.24 ) set
the pumps to pump a combined
(if using both pumps) 300 ul/min.Set pump
“B’ to 5% ACN and the ramp to 1 min.Starting ramp will start pumps.
Pump the first five min. into
waste to prevent salts and
contaminates entering MS.
Note:Always
increase pressure on column and pumps
slowly by ramping over time to prevent pressure
surges on columns which can
create dead space at top of column which will
decrease quality of separation.
Note:All
material going through column ends up in
MS ion detector.It is important not to
add contaminates or too much material to the MS
or the only ions that will be
seen will be the contamination.It is
always good to start with very small amount of
sample and increase if needed.
Run solvent through the column
to equilibrate it for at
least 6 to 10 volumes of solvent (5-10 min.).Switch from waste to load and monitor the
ions.The
ions of interest can be monitored
individually by typing their mass numbers (M +1)
in the middle screen, one at a
time.
Set up a full scan program.Open a method file and save as your group
and date under 3590.Set pump to 5% B at start, increase to 15%
B
over 5 min., then increase to 30% B over 2 min.,
return to 5% B over 1 min and
keep at 5% for 4 min. Set acquisition time to 12
min.
To inject sample (filter
samples before injection) from
auto-sampler you can use quick inject if you are
only injecting one sample or
you can select the sample table screen and set the
parameters in this screen,
including sample name, type, volume
injected, and auto-samplernumber.
.
Inject 10 ul of one of
standards (20 ug/ml) to determine the
retention times of acetaminophen, caffeine,
acetylsalicylic acid and ibuprofen.
Start the run by clicking the begin icon at the very bottom of the
sample window.
After completion of the first
run, the retention times of
the various constituents in your sample are known.
Make a combined standard of the
four standards, acetaminophen,acetylsalicylic
acid, caffeine, and
ibuprofen at concentrations
ranging from 5 to 100ug/ml (use
diluting solution).
Week 2:
Using the retention times from
week 1, retention time
segments (in method window) can be chosen to
correspond with the retention time
of each ion.Each time segment can have
its own set of conditions to optimize collection
of the ion of interest, and/or
to fragment the ion to product ions.Set
the CID voltage(s) to produce the 1st
(most abundant) product ions
for each precursor ion.
The conditions are set by
making a new method and selecting
MS-MS mode and setting the various parameters
under MS/MS parameters and the
sub windows.You will need to set the
capillary voltage, CID voltage, % RF, drying gas
pressure (35 p.s.i.)) and
temperature (keep at 350oC) and
nebulizer gas pressure (50 p.s.i.)
values for each time segment.
Rerun your standards and
samples, tylenol, ASA and the
mixture using your new MS/MS method with the time
segment information added to
your method.In this run you will get the
major product ions in each time segment.
Integrate a product ion of
standards and sample (hold I key
down and use cursor to draw baseline) to get area
related to concentration.You can
also integrate using the integration icon
in the tool bar.
Identify compounds by product
ions.
Compare the one standard run at
full scale, to it’s product
ion, to determine if sensitivity and/or
selectivity is better with product ion
(transition) or full scale precursor ion
Draw standard curve in “Excel”
and determine concentration
of medication.
See if label agrees with your
analysis.
Note:Make
sure the material going into the needle
of the ion spray is always washed out with
solvent before the pumps are shut
off or syringe pump is stopped or the needle
will plug and have to be replaced.
Data:
Identify each medication by
retention time and product ions.
Identify the fragmentation
patterns of the original and
product ions and compare to literature patterns.
Determine concentration of
medication and compare to claimed
concentration.
Discussion:
Discuss your results in detail.
Discuss the methodology,
improvements that are possible.Discuss
the fragmentation patterns and what is occurring
using literature sources.Discuss
concentration results, how it
compares with label and reasons which might affect
results, etc., etc.
Final
Questions:
What are the reasons
to use LC-MS versus GC-MS for the analysis of
pharmaceutical and biological samples.
What are some of the
problems associated with the use of this
instrument for quantitative analysis?Compare
advantages (disadvantages) of using full scan
ions verses product ion from transitions for
quantitation,and identification.
What parameters would
you have to control to produce a library capable
of identifying compounds based on their
fragmentation pattern?
The instrument for
this experiment is set to positive mode (can
only see positively charged ions).What
is the other routine mode the instrument can be
set in and when would this mode be used and why?