The
MS in MCAL uses electro-spray
ionization (ESI) and Ion Trap technology to
detect sample ions. The analyte
ions are nebulized and dried (and become charged
ions).The
ions can be scanned in full scale mode,
MS-MS mode or can be further fragmented and the
product ions scanned. The MS is
connected to a binary LC system.
Laboratory Week 2:
HPLC-MS
Prior to using the
LC-MS, there are a few important points to
check:
·Ensure the Solvent bottle has
significant levels: 30% Acetonitrile.Both
pump lines should be in this bottle.
·Ensure the source on the Mass
spec is connected to the LC via the red tubing rather
than the syringe pump
·Equilibrate the column for at
least 5 min
·Always slowly increase flow
rate (ramping over 1-2 min) as to avoid sudden
pressure surges to the system
·Make sure you wash the
autosampler before injecting your first sample
(ask the instructor to show you
how, it just needs to be done before you start
your samples).
Part A: Start up
LC-MS
1.At the top of the system
control software; select window in
the
tool bar and select 212 LC to open the pump
control window.There will be a graph display plotting
each
pump’s pressure as well as solvent.
2.Click the manual control button
and a new window will pop up.Enter
0%
under pump b and a flow rate of 250 uL. Ramp
time 1 min.Click start and the pumps should start
and
begin recording on the graph.
3.Monitor the pressure.It
should be fairly stable. Slight
fluctuations are normal; pay attention to the
scale.If
it is fluctuating more than 30 p.s.i.
there may be a problem such as an air bubble.Consult the instructor if this is the
case.
Part B: Prepare
Solutions
Stock
standards (dissolved in
M-Q water):
Caffeine
(1 mg/ml)
Theophylline
(1mg/ml).
30%
Acetonitrile + 0.1% Formic Acid
Standards for caffeine
determination with theophylline as internal
standard.
The
easiest way to prepare
standards is serial dilution, but remember you
need 1.0mL of each standard to
put in the HPLC vial for analysis. Also, to do
serial dilution you will need to
use Theophylline as the diluent in order to keep
the concentration constant.
Make a 10 mL solution of 80 ug/mLTheophylline
and 2 mL solution of of 160 ug/ml both usingthe
stock and 30% Acetonitrile/formic Acid.
Prepare the following Standards
by serial dilution ( using Theophyline as
diluent)
a. 80
ug/mL
Caffeine.
Theophylline 40
ug/ml
b. Caffeine:
10 ug/ml
Theophylline 40
ug/ml
c. Caffeine:
20 ug/ml
Theophylline 40
ug/ml
d. Caffeine:
40 ug/ml
Theophylline
40
ug/ml
e. Caffeine:
80 ug/ml
Theophylline 40
ug/ml
Caffeine
Unknowns:
·Cola, Diet
Cola, diluted 1 to 5 and 1 in 10
·Energy
drinks 1 in 10 and 1 in 20
All
your unknowns should be diluted
in 30% acetonitrile, 0.1% formic acid.Addtheophylline to each of the caffeine unknowns
so
that the final concentration is 40
ug/ml. This is your internal
standard.
Part C: Set up
Method and Run Standards and samples
Set up an LC-MS
method:
1.Open the method builder located on
the tool bar at the top right of
the screen, it is the second from the top.Click File then New Method
and save it in the 3590 2016
folder. Click next three times and then finish
and you will be able to save
your method.
2.On the left is a navigation pane of
all components of the
method.Click
on pump program and set
the following pump program in the window on the
right:
0-12min:
100% A
The flow rate should be set to 250
uL/min .
3.Double Click the Mass Spec
Aqcuisiton
in the navigation window:
a.Make sure Scan Type is full
b.Under Active
Segment name the
segment and set the segment end time.
c.Under Electrospray
Ionization Parameters, set the nebulizing pressure to 50 and
the drying gas pressure to
30.Drying
gas temp should remain at
350.
d.Under the Full Scan
parameters, set the
Capillary voltage to 70, the start and
end mass to 100 and 400 respectively, and the RF
loading to 85%.
4.Save your method and upload. To
upload, go to the System Control
window and there is a button at the top
displaying the active method.Left
click and select Activate Method. Find
yours and open it and it will activate.
5.Under the 500 MS window;
select File then New Sample list and save it in
the 3590 2016 folder.
a.Enternames of standards and your unknowns; the
correct vial # position and an injection volume
of 10 uL.
b.Click Data Files in the bottom right
hand corner and make sure the data
is allocated to your folder.
c.Click Begin.Observe
the tool bar.After a minute or so; you should see the
NOT READY
change to WAITING
and you will hear the
autosampler start up.
Part D: Data Analysis
1.Open the MS data review
software (found on the same toolbar as the
method editor)
2.Locate your chromatogram from
your folder on the left hand navigation window
and click on it.
3.Use the curser and click on
the peak of interest in the TIC; this will open
the RIC in a window below and
display the masses associated with that
chromatographic peak.
4.Integrate the area ofeach
of your peaks in your standard and
unknowns to create a curve in excel and
determine the concentration of your
unknown:
a.Extract the molecular ion by
typing in the window at the bottow left of the
TIC.This
will filter out only the molecular ion
from the rest of the ions through this peak.
b.Locate the set single click action button on
the tool bar above the TIC and
set it to integrate.
c.Click the mouse cursor where
you want to start the integration and drag it to
where you want to end.The
peak will fill in color and provide an
area and RT at the top.If you
can’t see
it; you will need to zoom in or out by resetting
the single click action back
to zoom.
d.Record the area
e.Repeat for all peaks in all
samples.
Refer to original
lab report for Data Analysis procedures –
Internal and External Standard.
Once samples are
finished; move the inspirator for
pump B into 100% Acetonitrile and change theto 80 % in the pump manual control
window.Run
for 10 minutes to clean out the
column.
Once finished; you
need to turn off the pumps from the 212 pump
window by clicking Stop
Pumps. Also
turn of the trap clicking on it in the
schematic; then turn the ONLY the ion
source back on.